143 ELECTROSURGICAL SAFETY I-04_ElectroSafety-25_110124.indd 143 I-04_ElectroSafety-25_110124.indd 143 11/7/2024 10:34:32 AM 11/7/2024 10:34:32 AM

144 ELECTROSURGICAL SAFETY TABLE OF CONTENTS 1. Injury Prevention ............................................................................................................................... 146 Fire prevention, patient and personnel injuries 2. ESU Generator .................................................................................................................................... 148 Alarms, power settings, maintenance, documentation 3. ESU Accessories P ............................................................................................................................... 150 Cords, insulation integrity, active electrode, dispersive electrode 4. Minimally Invasive Surgery .............................................................................................................. 154 Conductive trocar systems, active electrode monitoring 5. Implanted Electronic Devices P ...................................................................................................... 154 Cardiac implants, cochlear implants 6. Argon-Enhanced Coagulation ........................................................................................................... 158 7. Education P ........................................................................................................................................ 158 MEDICAL ABBREVIATIONS & ACRONYMS ESU – Electrosurgical unit AEC – Argon-enhanced coagulation EMI – Electromagnetic interference IED – Implanted electromechanical device MAUDE – Manufacturer and User Facility Device Experience SSEP – Somatosensory evoked potentials IFU – Instructions for use FDA – US Food and Drug Administration CIED – Cardiac implanted electronic device ID – Identification P indicates a recommendation or evidence relevant to pediatric care. I-04_ElectroSafety-25_110124.indd 144 I-04_ElectroSafety-25_110124.indd 144 11/7/2024 10:34:32 AM 11/7/2024 10:34:32 AM

ELECTROSURGICAL SAFETY 145 T he Guideline for Electrosurgical Safety was approved by the AORN Guidelines Advisory Board and became effective as of July 29, 2020. Information about the systematic review supporting this guideline, including a description of the evidence review, the PRISMA flow dia- gram, the evidence rating model, and the evidence summary table is available at https://www.aorn.org/evidencetables. Purpose This document provides guidance to the perioperative team for the safe use of electrosurgical units (ESUs), electrocautery devices, and argon-enhanced coagulators. An ESU includes a generator, an electrical cord and plug, and accessories. The accessories include the active electrode with tip(s), dispersive electrode, foot switch with cord (if applicable), adapters, and connectors.1 The electrosurgical generator performs three functions. First, it converts the low-frequency alternating current received from the electrical circuit in the wall, which is at 60 Hertz, to approximately 500,000 Hertz (ie, radio frequency). The second function enables adjustment of the power setting. The third function controls the propor- tion of time over which a waveform is produced (ie, the duty cycle). These waveforms are known as cut, which is heating of the cellular water that leads to the cell burst- ing; coagulation, which causes a temperature rise in the cells leading to cellular dehydration and shrinkage; and blend, which is a modulated form of cut that results in an output with a higher voltage than cut at the same power setting.1,2 The radio-frequency energy produced is trans- ferred to the patient by various modalities, including monopolar, bipolar, advanced bipolar, bipolar ligating- cutting, and tripolar (ie, plasma knife) devices and argon- enhanced coagulation (AEC). These modalities are used to cut, coagulate, dissect, ablate, and shrink tissue. The monopolar modality transfers the energy to the patient through an active electrode that usually has only a single tip. The intended current flow is from the generator through the active electrode cord to the active electrode tip, through the patient to the dispersive electrode, and then through the dispersive electrode cord back to the generator. The monopolar modality requires both an active and a dis- persive electrode.1,2 The bipolar modality has a two-tip electrode that trans- fers the energy to the patient. The current pathway for this device goes from the generator through the cord to one tip of the forceps, through the tissue between the tips to the other tip, and then back to the generator. The current flows only through the tissue that is between the forceps tips. The only accessory required is the active electrode and the asso- ciated cord. Bipolar electrosurgery can be used as an alternative to monopolar electrosurgery when devices or implants would be in the current pathway between the monopolar active and dispersive electrodes.1,2 The advanced bipolar modality uses bipolar electrosurgery with a computer-controlled tissue feedback response system to sense tissue impedance. This allows for the continuous adjustment of the voltage and current generated by the unit. Continuous adjustment permits use of the lowest possible power setting that will achieve the desired tissue effect. The advanced bipolar modality does not require a dispersive elec- trode and requires less voltage. The energy flows only through the tissue that is between the forceps tips.1,2 Some of the devices also have a cutting mechanism, allowing for cut- ting and coagulating of tissue between the forceps tips.3 The tripolar device has a tripolar tip consisting of a central pole and an outer pole on either side. The current flows from the center pole to the outer poles, creating a corona of energy that makes a blade-like incision. The pri- mary current alternates with a second current that passes from one of the outer poles to the other, resulting in simul- taneous cutting and coagulation.1,3 Argon-enhanced coagulation, also known as argon beam coagulation, is radio-frequency coagulation from an electro- surgical generator that delivers monopolar current through a flow of ionized argon gas. The risks related to this modal- ity are similar to those for monopolar electrosurgery with the addition of the risk for gas emboli.4 The electrical current used by the ESU consistently flows on a pathway from the wall, through the device, to the accessories, through tissue, through the accessories, back to the device, and to the wall. Most adverse events associated with electrosurgery are related to the current trying to flow back to ground.1,3 The majority of adverse events that result from the use of electrosurgery or electrocautery are burns. Burns may be caused by direct application that results in thermal spread beyond the intended target tissue, insulation failure, antenna coupling, direct coupling, capacitive coupling, residual heat, or inadvertent activation and may be described as alternate site injuries.1,5-12 Adverse events have been reported to occur during various procedures and with all sur- gical approaches.13-29 Burns may occur anywhere the tip of the active electrode is placed, under the dispersive electrode, inside the body during laparoscopic procedures, or at trocar sites.1,5-12 Other types of adverse events include electrical shocks, electromagnetic interference (EMI), and fires.1 2025 Guidelines for Perioperative Practice Last revised: July 2020. Copyright © 2025 AORN, Inc. All rights reserved. GUIDELINE FOR ELECTROSURGICAL SAFETY I-04_ElectroSafety-25_110124.indd 145 I-04_ElectroSafety-25_110124.indd 145 11/7/2024 10:34:33 AM 11/7/2024 10:34:33 AM

146 ELECTROSURGICAL SAFETY Electrical shocks do not often involve the patient because they occur off the sterile field; if strong enough, however, a shock can cause injury to a person in contact with the device or the electrical cord. A shock may result in a burn, another type of injury, or no injury.1 Electromagnetic interference may cause oversensing, cause a function reset, or stop the functioning of an implanted electromechanical device (IED) and could also damage a lead or the pulse generator itself.30,31 In a retro- spective study of 1,398 patients undergoing cardiac pace- maker generator replacement or upgrade surgery, Lin et al32 found that four patients (0.3%) developed output failure or an inappropriate low pacing rate during use of electro- surgery. The researchers also reviewed the Manufacturer and User Facility Device Experience (MAUDE) database for events or patient injuries related to pacemaker malfunc- tion and found 37 cases of pacemaker malfunction related to the use of electrosurgery. The adverse events included the pacemaker resorting to the backup pacing mode (32.4%), loss of output or capture function (59.5%), inap- propriate low pacing rate (5.4%), and ventricular fibrilla- tion (2.7%). The dates of review of the MAUDE database were not included in the report. Other authors have also reported on situations in which EMI during the use of electrosurgery led to complications, including ventricular tachycardia,33 ventricular fibrillation,28,34,35 and inappropri- ate defibrillation that occurred when an implantable cardio- verter defibrillator sensed EMI from an electrosurgical active electrode as ventricular fibrillation.36-38 The active electrode is the most common ignition source in OR fires.1 A fire may or may not result in patient burns. Electrosurgical accessories, including the active and dispersive electrodes, have been implicated as the cause of injuries.3,15,20,39,40 Overbey et al16 reviewed reports to the MAUDE database from January 1, 1994, to December 31, 2013, and found 3,553 injuries and 178 deaths related to energy-generating devices. The energy-generating devices were classified as monopolar instruments (n = 1,670; 44.8%), radio-frequency/microwave ablation devices (n = 728, 19.5%), advanced bipolar devices (n = 538, 14.4%), ultra- sonic devices (n = 350, 9.3%), bipolar instruments (n = 270, 7.2%), plasma beam monopolar devices (n = 163, 4.4%), and other (n = 12, 0.3%). The incidents involved thermal burns (n = 2,353, 63.1%), hemorrhage (n = 642, 17.2%), mechanical fail- ure (n = 442, 11.8%), and fire (n = 294, 7.9%). The authors concluded that the risk for injury from surgical energy- generating devices is significant and warrants further research and education. A limitation of the evidence is that randomized con- trolled trials related to electrosurgical injury prevention may expose patients to harm and, as such, would not be eth- ical. A limited number of other types of studies have con- tributed valuable knowledge to the field. However, interpre- tation of these studies is limited by the nature of this type of research, which can only show association among study variables and cannot determine causation. Because of a lack of research on interventions to prevent injury from the use of ESUs, much of the available evidence is based on gener- ally accepted practices and expert opinion. The following subjects are outside the scope of this guideline: • • general fire safety (See the AORN Guideline for a Safe Environment of Care)41; • • surgical smoke safety (See the AORN Guideline for Surgi- cal Smoke Safety)42; • • selection of endoscopic distention fluid (See the AORN Guideline for Minimally Invasive Surgery)43; • • procedure-related decisions (eg, the amount of time the tissue is exposed to the active electrode); • • therapeutic diathermy; • • use of electrical dental equipment (eg, battery-operated curing lights, ultrasonic baths, ultrasonic scalers, electric pulp testers, electric toothbrushes); and • • selection of electrosurgical devices. 1. Injury Prevention 1.1 Assess the patient preoperatively for the pres- ence of foreign bodies (eg, IED, jewelry, pros- thetic implants). [Recommendation] Assessment will help to determine the presence of items that will create the need to take actions to prevent injury. 1.2 Place patient monitoring electrodes (eg, electro- cardiogram, oximetry, fetal) as far as possible from the surgical site.44 [Recommendation] Placing electrodes as far as possible from the sur- gical site decreases the risk for a burn at the elec- trode site.44 1.3 Use alternate technologies (eg, bipolar, ultra- sonic) instead of monopolar electrosurgery when neuromonitoring electrodes (eg, somatosensory evoked potentials [SSEP]) are present (See Section 5. Implanted Electronic Devices).45 [Recommendation] The findings of a laboratory study conducted by Townsend et al45 support the use of alternate tech- nologies. The researchers examined the tempera- ture of the surface of a porcine tissue model onto which a neuromuscular monitoring lead was placed. They found an insignificant increase in the temper- ature at the neuromonitoring site with the use of an advanced bipolar device (0.2° C ± 0.4° C [0.4° F ± 0.7° F]) and an ultrasonic device (0° C ± 0.3° C [0° F ± 0.5° F]) compared to the temperature increase with the use of a monopolar ESU (39° C ± 13° C [70° F ± 23° F]) or an argon beam generator (34° C ± 15° C [61° F ± 27° F]). The researchers recommended using alternative technologies when neuromonitoring electrodes are used. I-04_ElectroSafety-25_110124.indd 146 I-04_ElectroSafety-25_110124.indd 146 11/7/2024 10:34:33 AM 11/7/2024 10:34:33 AM

ELECTROSURGICAL SAFETY 147 1.4 During the use of monopolar devices, prevent con- tact between the patient and metal objects.44,46-48 [Recommendation] Alternate site burns may occur because the electricity seeks an alternate pathway to ground, from the patient’s skin to the metal object, instead of flowing to the dispersive electrode.44,46-48 Low- quality evidence indicates that the risk for alternate site burns related to the phenomenon of stray cur- rent leakage has been almost eliminated by techno- logical improvements in ESUs.49,50 A clinical guideline supports preventing contact between the patient and metal objects (eg, OR beds, stirrups, positioning devices, safety strap buckles) to prevent alternate site burns.51 1.5 Remove all metal jewelry that will be between the active and dispersive electrodes.25,49,50,52 [Recommendation] Metal jewelry, including body piercings, subdermal implants, and transdermal or microdermal implants, some of which may not be removable, presents a risk for thermal injury at the location where the jewelry touches the skin. The injury is caused by the heating of the metal as the electrical current moves through it.52 Nguyen et al53 conducted a bench study to exam- ine the effect of monopolar radio-frequency energy on the temperature of bovine myocardial tissue surrounding implanted metal objects. The objects consisted of insulated and noninsulated esophageal temperature probes, copper wire, a defibrillator lead, and a circular mapping catheter. The researchers applied radio-frequency energy to the tissue at 30 W using an ablation catheter. They found the temperature in the tissue surrounding the objects was elevated by a mean of 5.8° C to 8.9° C (10.4° F to 16° F) when the objects were between the active and dispersive electrodes and at a dis- tance of 1 mm from the site of the active electrode. The researchers concluded that a significant rise in myocardial tissue temperature occurs when radio- frequency ablation is performed near metal items. In contrast, the results of a quasi-experimental study conducted by Sheldon et al54 did not support the need to remove microdermal implants. The researchers measured the skin temperature at microdermal implant sites located between the active and dispersive electrodes on swine skin. The disper- sive electrode was placed 2 cm below an implant made of surgical stainless steel. The simulated surgical site was 2 cm above the implant. The researchers set the monopolar electrocautery at 30 W coagulation and administered it continuously for 30 seconds. They used thermal imaging to measure the surface skin temperature and assessed the tissue using gross examination and a histological evaluation. The researchers found the skin temperature at the surgical site rose 27.7° C (49.9° F), and at the implant site the temperature was elevated 1.6° C (2.9° F). The researchers also measured the tempera- ture at a control site on the contralateral side. The temperature increase at the subdermal site was 2° C (3.6° F) when the energy was applied on the contra- lateral site, which was not a significant difference. The researchers found no thermal injury on the tis- sue in close proximity to the subdermal implant site. They concluded that subdermal implants do not have to be removed even when in close proximity to the active electrode.54 1.5.1 When jewelry cannot be removed, • • notify the surgeon, • • consider using alternate technologies if jewelry is in the current pathway,25,49,50,52 • • provide education to the patient regarding potential adverse events and document the education provided, • • assess all jewelry sites postoperatively for any evidence of burns, and • • document preoperative and postoperative assessment of jewelry sites. [Recommendation] 1.6 Follow the manufacturer’s written instructions for use (IFU) for all ESU components.46 [Recommendation] 1.7 Perform the following fire prevention interven- tions when using an electrosurgical device: • • use technologies other than monopolar (eg, bipolar devices, coblation technology, non- energy-applying instruments) during surgical procedures on anatomical structures that present special fire hazards (eg, bowel, trachea)12,19,23,26,55-59; • • use moist radiopaque sponges near the igni- tion source12-14,19,21,57,60,61; • • remove alcohol-soaked sponges or other appli- cation devices from the sterile field14,18,22,62-65; and • • allow alcohol-based solutions to dry and fumes to dissipate before using any type of electrosurgical active electrodes.14,18,22,62-65 [Recommendation] Low-quality evidence12-14,17,19,21-23,26,55-58,60-65 and a clin- ical guideline66 identify electrosurgical devices as an ignition source and support performing these inter- ventions. For more information, see the fire pre- vention recommendations in the AORN Guideline for a Safe Environment of Care.41 Matt and Cottee56 conducted a descriptive labo- ratory study in which they placed a bipolar radio- frequency ablation device in direct contact with I-04_ElectroSafety-25_110124.indd 147 I-04_ElectroSafety-25_110124.indd 147 11/7/2024 10:34:33 AM 11/7/2024 10:34:33 AM

148 ELECTROSURGICAL SAFETY flammable materials found in the OR: alcohol, four individual salt packets dissolved in alcohol, two pumps of ethanol-based foam hand sanitizer, saline- dampened gauze, plastic packaging filled with 5 mL of saline, and a 4.0 endotracheal tube with 100% oxy- gen flowing through the tube. The researchers found that the bipolar radio-frequency ablation device did not ignite the various materials even when set at the highest settings (ie, 9 ablation, 5 coagulation). They concluded that the tested bipolar radio-frequency ablation device did not produce an adequate amount of heat to serve as an ignition source at the surgical site, and therefore decreased the potential for fire. In a laboratory study, Roy and Smith55 compared the time to ignition of a chicken carcass (as a human tissue surrogate) when using a standard monopolar device at 15 W coagulate mode versus a bipolar radio-frequency ablation wand set at 9, 7, and 3 in the ablate mode. The researchers inserted an endo- tracheal tube into the gut of the chicken, with 100% oxygen flowing through the tube. Contact between the chicken and the device was made near the end of the endotracheal tube. Ignition occurred after 25 to 80 seconds during use of the monopolar device, and no ignition occurred during use of the bipolar device. The researchers concluded that use of a bipo- lar radio-frequency ablation device eliminates the risk for fire during open cavity surgery. 1.8 Perform the following fire prevention interven- tions when using an electrocautery device: • • use radiopaque sponges moistened with non- flammable liquid near the ignition source,67 • • apply the protective cap when the cautery is not in use,68 and • • remove the wire loop and batteries from the electrocautery device before discarding it.59,68 [Recommendation] An electrocautery device has been identified as an ignition source.68 In a non-experimental study, Axelrod et al67 tested the flammability of plastic tubing, an OR cap, paper tape, dry and moist cotton gauze, a paper gown, dry and moist cotton towels, a plastic drape, and a cellu- lose eye spear. The researchers found all the items except the moist towel and moist cotton gauze would ignite, melt, or char when touched with an electrocautery device at 1.5 or 3.0 volts. The researchers recommended using moist sponges near the site of electrocautery use. 1.9 When patient or personnel injuries or equipment failures occur during the use of an electrosurgi- cal generator or accessories, • • remove the electrosurgical generator and accessories from service; • • retain all accessories and packaging if possi- ble69; and • • report the adverse event details, including device identification and maintenance and service information, according to the health care organization’s policy and procedures. [Recommendation] A clinical guideline supports removing malfunc- tioning ESUs from service.51 Retaining the generator, accessories, and packaging can facilitate a complete incident investigation.69 1.9.1 The health care organization must report to the US Food and Drug Administration (FDA) and the manufacturer when a medical device is suspected of having contributed to a death, or only to the FDA if the medical device manu- facturer is unknown.70 [Regulatory Requirement] The FDA provides specific guidance for man- datory reporting.70 1.9.2 The health care organization or personnel may report ESU malfunctions to the FDA.70 [Conditional Recommendation] The FDA provides specific guidance for vol- untary reporting.70 1.9.3 The health care organization must report to other regulatory bodies as defined by state or local regulations. [Regulatory Requirement] 2. ESU Generator 2.1 Keep safety and warning alarms and activation indicators on the generator operational, audible, and visible at all times.1,44,69 [Recommendation] Safety and warning alarms alert the operator to potential equipment failure. The indicators and alarms immediately alert the perioperative team when the active electrode is activated.1,69 2.2 Select the lowest power setting on the electro- surgical generator that achieves the desired result.45,71,72 [Recommendation] High-quality evidence supports that using the lowest power setting needed to achieve the desired result reduces • • damage from capacitive coupling,71 • • the amount of undesirable tissue effect (ie, depth of tissue debridement plus depth of chondrocyte death in underlying cartilage),73 • • the size of the ablation area,74 • • the depth and area of injury,75 • • the amount and depth of cell death,76 • • damage from antenna coupling,72 I-04_ElectroSafety-25_110124.indd 148 I-04_ElectroSafety-25_110124.indd 148 11/7/2024 10:34:33 AM 11/7/2024 10:34:33 AM

ELECTROSURGICAL SAFETY 149 • • the temperature of the skin under a neuromuscu- lar and an electrocardiogram lead,45 • • thermal spread,77 and • • development of esophageal ulcerations.78 These studies were quasi-experimental, and all of the studies except one78 were conducted in labora- tory settings using animal tissue.45,71-77 Lower-quality evidence also showed that using the lowest power setting needed to achieve the desired effect reduced the incidence of • • thermal spread,79,80 • • antenna coupling,81 • • thermal injury,82 • • hemorrhage,83 and • • EMI.84,85 These studies were nonexperimental, and all of the studies except one83 were conducted in laboratory settings using animal tissue.79-82,84,85 Clinical guidelines also support use of the lowest effective power setting.59,86 However, the findings of a nonexperimental study by Metzner et al87 conflict with those of the previous studies. The researchers assigned 30 patients undergoing pulmonary vein iso- lation with an endoscopic ablation system to one of three groups: • • in Group 1 (n = 10), ablation was performed at power settings of 5.5 W and 7.0 W; • • in Group 2 (n = 10), ablation was performed at power settings of 7.0 W and 8.5 W; and • • in Group 3 (n = 10), ablation was performed at power settings of 8.5 W and 10.0 W. Pulmonary vein isolation was successfully per- formed in each of the groups (ie, 69% in Group 1, 73% in Group 2, 90% in Group 3). One patient in each of Groups 2 and 3 developed esophageal thermal lesions; none were reported in Group 1. The researchers concluded that higher energy settings increase efficacy and are safe to use. They also stated that the study was small and further research should be conducted before final conclusions are drawn.87 2.2.1 Confirm the power settings on the electro- surgical generator with the operator before activation.51 [Recommendation] Clinical guidelines support confirming the power settings before activation.51,86 2.2.2 If the operator repeatedly requests an increase in power, confirm the integrity of the entire circuit including the generator and all accessories.69,88,89 [Recommendation] A nonintact circuit, such as a loss of adhesion between the dispersive electrode and the patient, may cause ineffective coagulation and cutting, resulting in a request by the user for an increase in power.88 2.3 Mount the electrosurgical generator securely to a tip-resistant cart or shelf.51 [Recommendation] A clinical guideline supports mounting the gener- ator on a tip-resistant cart or shelf to help prevent the unit from falling to the floor and possibly caus- ing injury to the patient or personnel.51 2.4 Do not place items including equipment or con- tainers of liquids on the electrosurgical genera- tor.69,90,91 [Recommendation] Use of the generator as a table may create a fire hazard or cause malfunctioning of the equipment placed on top of the generator.69,90 Putting containers of liquids on the generator may increase the poten- tial of an unintentional activation, a device failure, or an electrical hazard resulting from liquids enter- ing the electrosurgical generator.69,91 A clinical guideline advises against putting con- tainers of liquids on the generator.51 2.5 Qualified personnel (eg, a biomedical engineering services representative) must perform periodic monitoring, inspection, testing, and maintenance on the electrosurgical generator.92,93 [Regulatory Requirement] The Centers for Medicare & Medicaid Services requires that equipment be inspected, tested, and maintained.92,93 2.5.1 Document inspection and maintenance according to the health care organization’s policy. [Recommendation] 2.6 Document the following in the patient’s medical record in a manner consistent with the health care organization’s policies and procedures: • • electrosurgical generator identification (eg, serial or biomedical number), • • location of dispersive electrode placement, and • • the patient’s skin condition before applica- tion and after removal of the dispersive electrode.94 [Recommendation] 2.7 Follow the electrosurgical generator and acces- sory manufacturers’ IFU when multiple ESUs are used simultaneously.69 [Recommendation] A clinical guideline recommends following the manufacturers’ IFU.59 2.7.1 Label accessories to correspond with the generator to which they are attached.95 [Recommendation] Labeling individual generators and accesso- ries to match helps identify which ESU requires adjustment or, during a move by the operator I-04_ElectroSafety-25_110124.indd 149 I-04_ElectroSafety-25_110124.indd 149 11/7/2024 10:34:33 AM 11/7/2024 10:34:33 AM

150 ELECTROSURGICAL SAFETY from one side of the table to the other, which operator is using which unit.95 2.7.2 Use one single-use dispersive electrode for each generator when multiple generators are used and • • place the dispersive electrodes as close as possible to the surgical site, • • do not overlap the dispersive electrodes, and • • place the dispersive electrodes equidistant from the surgical site when it is a single site.39 [Recommendation] Equidistant placement results in the equal distribution of the electrical current.39 3. ESU Accessories 3.1 Use accessories that are compatible with the gen- erator as specified in the manufacturers’ IFU. [Recommendation] A clinical guideline supports using only compati- ble accessories.86 3.2 Place the cords that connect the monopolar active electrode and the dispersive electrode to the gen- erator as far as possible from or perpendicular to other cords (eg, cardiac and neuromuscular moni- toring cords, light cords, second electrosurgery cords, camera cord).45,72,81,96-98 [Recommendation] High-quality evidence supports placing these cords as far as possible from or perpendicular to other cords.45,98 Robinson et al98 conducted a randomized con- trolled trial that included 84 patients undergoing laparoscopic cholecystectomy. The researchers mea- sured the amount of thermal injury at the skin adja- cent to the umbilical trocar site in control group procedures (n = 42) during which active electrodes and camera cords were oriented parallel and located in close proximity and in experimental group proce- dures (n = 42) during which the cords were sepa- rated. The researchers found that thermal injury was significantly less in the experimental group than in the control group (31% versus 57%). They concluded that cords should be separated and not oriented par- allel or located in close proximity. Townsend et al45 used a porcine model to examine the temperature of the surface of the tissue where a neuromuscular and an electrocardiogram lead were attached. The items were tested with an ESU power setting at 30 W and 15 W and cords placed close to or 15 cm away from the electrodes. The researchers noted a significant decrease in the temperature when the lower power setting was used and when the cords were not in close proximity. Based on these findings, they recommended setting the power as low as possible and placing the wires as far from the electrodes as possible. Lower-quality evidence also supports placing cords that connect the monopolar active or disper- sive electrode to the generator as far as possible from or perpendicular to other cords.72,81,96,97 Jones et al81 illustrated antenna coupling in a labo- ratory study. The researchers measured the tissue temperature in a bovine liver using an L-hook device as the active electrode and an unlit 10-mm laparo- scope as the instrument receiving the energy. The researchers compared the temperature at the end of the scope when the cords were parallel and bundled to the temperature when the cords were completely separated. They found the increase was greater when the cords were bundled than when the cords were separated (38.2° C [68.8° F] versus 15.7° C [28.3° F]). The researchers recommended separating the camera and light cords and the wires leading from the genera- tor to the active electrode. Robinson et al72 found simi- lar results and drew the same conclusion in a descrip- tive laboratory study using bovine muscle tissue. In a non-experimental laboratory study, Townsend et al97 measured the temperature at the end of a 10-mm 30-degree laparoscopic telescope and a Maryland lapa- roscopic grasper after activating a monopolar L-hook placed 4 cm away during simulated four-port laparo- scopic procedures on porcine liver tissue. The ESU was activated 10 times with the monopolar generator set at 30 W coagulation mode. The researchers found a signif- icant increase in temperature when the active elec- trode and the telescope cord were in close proximity compared to when the cords were separated at the end of the telescope, which has a cord, but not at the end of the Maryland grasper. Townsend et al96 achieved simi- lar results when the dispersive electrode cord was in close proximity to the camera cord compared to when they were separated. 3.3 Secure electrosurgical accessory cords to the sterile drapes • • with a non-piercing plastic or other noncon- ductive device, and • • in a manner that does not crush or damage the cord or compromise the sterile field.69 [Recommendation] Plastic devices do not conduct electricity if there is leakage. Use of a non-piercing device decreases the potential for sterile field contamination that could be caused by holes made in the drapes. Crush- ing or damaging a cord may create a place at which the current can escape from the cord, creating a fire on the drapes or causing the device to malfunction.69 I-04_ElectroSafety-25_110124.indd 150 I-04_ElectroSafety-25_110124.indd 150 11/7/2024 10:34:33 AM 11/7/2024 10:34:33 AM

ELECTROSURGICAL SAFETY 151 3.4 Track the number of uses of reusable accesso- ries (eg, cords, handles, shafts), and handle an accessory per the manufacturer’s IFU when the maximum number of uses has been reached. [Recommendation] Manufacturers’ IFU contain instructions on the maximum number of sterilization cycles for which integrity testing has been performed.86 3.5 Visually inspect electrosurgical accessories for dam- age (eg, insulation breakage) before and after use and, if reusable, during reprocessing.8,20,24,29,48,69,87,99-101 [Recommendation] Inspection helps to identify malfunctioning acces- sories (eg, active electrode internal wires, electrode insulation, sheaths) that have caused procedure delays (eg, while replacement equipment is obtained), injury to personnel or patients, or ignition of surgical drapes or clinicians’ gowns.8,29,48,69,87,100,101 Refer to the AORN Guideline for Care and Cleaning of Surgical Instruments102 for more detailed recommen- dations for instrument inspection. Munro82 conducted a laboratory comparative study to identify mechanisms of thermal injury to a patient’s lower genital tract during radio-frequency resectoscopic surgery. The researcher used a tissue model to represent the uterus and created defects in the insulation on the electrode. He found injuries resulting from capacitive coupling occurred when the insulation was damaged. Montero et al103 examined 165 reusable laparo- scopic instruments for breaks in insulation and found 19% of the instruments had insulation failure. The researchers recommended that surgeons be made aware of the many insulation failures that can occur and adopt practices to decrease the potential for complications related to insulation failure. Clinical guidelines also support inspecting accessories.51,59,86 3.5.1 Use an active electrode insulation integrity tes- ter while preparing active electrodes for pack- aging during processing.104,105 [Recommendation] Moderate-quality evidence104,105 and a clinical guideline59 support the use of an active elec- trode insulation integrity tester. In a descriptive study, Tixier et al105 inspected and tested 489 instruments for insulation fail- ure. The researchers found 24.1% of the instru- ments failed a visual inspection and 37.2% failed with the use of an active electrode insulation integrity tester. Failure occurred in a median location in 50.4% of the laparoscopic instru- ments and in a distal location in 40.4% of the non-laparoscopic instruments. In another descriptive study, Espada et al104 tested 78 robotic and 298 laparoscopic instru- ments at 20 W and 2.64 kV in phase A and tested 60 robotic and 308 laparoscopic instruments at 20 W/1 kV and 20 W/4.2 kV in phase B. In phase A, the researchers found insulation failures in 80% of the robotic instruments and 36% of the laparoscopic instruments. In phase B, they detected failures in 81.7% of the robotic instru- ments and 19.5% of the laparoscopic instru- ments. The researchers recommended routine testing of laparoscopic and robotic instruments for insulation failure. 3.5.2 If an insulation failure is identified during pro- cessing, sterile processing personnel should remove the damaged device from service.86 [Recommendation] 3.6 An interdisciplinary team that includes mem- bers of the perioperative and sterile processing teams should determine a standardized commu- nication strategy and actions to perform when an insulation failure is found during processing. [Recommendation] 3.6.1 Include the following: • • the communication pathway between sterile processing and surgical personnel, • • physician notification, • • device handling, and • • documentation according to facility policy and procedure. [Recommendation] 3.7 The electrosurgical generator foot pedal activa- tion switch should be activated exclusively by the person in control of the active electrode.60,69 [Recommendation] Activation of the active electrode exclusively by the operator helps prevent unintentional activation of the device, minimizing the potential for patient and personnel injury.60,69 Clinical guidelines support activation of the active electrode exclusively by the operator.51,59 3.8 Place a fluid-resistant cover over the foot pedal when there is potential for fluid spills. [Recommendation] Covering the foot pedal with a fluid-resistant cover helps keep the foot pedal dry, therefore decreasing the potential for an electrical shock or a mechanical malfunction. Clinical guidelines support use of a fluid-resistant cover on the foot pedal.59,86 I-04_ElectroSafety-25_110124.indd 151 I-04_ElectroSafety-25_110124.indd 151 11/7/2024 10:34:33 AM 11/7/2024 10:34:33 AM

152 ELECTROSURGICAL SAFETY 3.9 When using an active electrode, • • connect the cord directly into a designated receptacle on the electrosurgical generator59,106; • • place the electrode in a clean, dry, noncon- ductive safety holster when it is not in use1,29,59,60,63,64,69,91; and • • use only adaptors approved by the manufac- turers of the ESU and the accessory. [Recommendation] A clinical guideline supports performing the listed interventions.86 Use of a nonconductive safety holster helps pre- vent the active electrode from falling off the sterile field and from unintentional activation.1,29,59,63,91 3.10 Do not activate the active electrode until it is in close proximity to the surface if fulgurating, or in contact with the target tissue if cutting or coagu- lating.1,8,40,47,48,107 [Recommendation] Activation of the active electrode only when it is in close proximity to or in contact with the target tissue helps prevent alternate site burns.1,8,40,48,107 Makedonov and Lee47 used a computer-generated model to determine the potential for an alternate site burn. They found that an alternate site burn would occur under following conditions: • • 20 seconds of continuous ESU activation, • • activation of the ESU without contact between the active electrode and the patient, • • a small alternate return site contact area of 1 cm2, and • • the use of a high-power (ie, 200-W) cut mode. The researchers concluded that to prevent alternate site burns, the active electrode should not be acti- vated prior to contact with the patient. A clinical guideline supports not activating the active electrode except when it is in close proximity to the target tissue.59 3.11 When using an active electrode tip, • • use the tip according to the manufacturer’s IFU; • • firmly seat the tip into the hand piece with no gaps between the tip insulation and the hand piece107; • • clean the tip • whenever there is visible eschar present, • away from the incision, • with a sponge moistened with sterile water or with an instrument wipe for nonstick- coated electrosurgical tips, and • with an abrasive electrode cleaning pad for noncoated electrodes29,52,69,106; and • • do not alter (eg, bend) the tip unless this is per- mitted by the manufacturer’s IFU. [Recommendation] Cleaning the tip decreases the amount of resistance; therefore the active electrode requires less power.29,52 3.11.1 An anti-stick solution may be used to decrease eschar buildup.108,109 [Conditional Recommendation] P High-quality evidence108,109 supports use of an anti-stick solution to reduce surgical time. In a ran- domized controlled trial, Baker and Ramadan108 measured the differences in the number of hand backs (ie, the number of times the electrode tip needed to be handed back to the scrub person for cleaning) and the length of surgery between pedi- atric adenoidectomy procedures in which an anti- stick phospholipid solution was (n = 31) or was not (n = 30) applied to the electrosurgery suction tip. The researchers found a 1-minute 45-second decrease in the length of the procedure and a mean decrease of three hand backs per procedure with use of the anti-stick solution. Roy et al109 found similar results in a randomized controlled trial of 50 patients undergoing rhytidectomy with use of a bipolar electrosurgical electrode. 3.12 Assess the patient’s skin before surgery in the vicinity of the potential dispersive electrode appli- cation site and after surgery at the point of contact with the dispersive electrode. [Recommendation] Preoperative and postoperative assessments are necessary to evaluate the patient’s skin for injuries at the dispersive electrode site. A clinical guideline supports performing a skin assessment.59 3.13 Use a single-use dispersive electrode according to the manufacturer’s IFU.49 [Recommendation] 3.13.1 Select the single-use dispersive electrode size based on the patient’s weight, as described in the manufacturer’s IFU (eg, neonate, infant, pediatric, adult).20,25,39,44,69 [Recommendation] P 3.13.2 Confirm that the manufacturer’s beyond- use date has not expired before opening the single-use dispersive electrode, and do not use it if the beyond-use date has passed. [Recommendation] 3.13.3 Open single-use dispersive electrode packag- ing immediately before the electrode is needed.49 [Recommendation] Opening the single-use dispersive electrode immediately before use helps decrease the potential for drying of the dispersive electrode’s adhesive and conductive gel.49 A clinical guideline recommends not opening the package until the electrode is needed.51 I-04_ElectroSafety-25_110124.indd 152 I-04_ElectroSafety-25_110124.indd 152 11/7/2024 10:34:33 AM 11/7/2024 10:34:33 AM

ELECTROSURGICAL SAFETY 153 3.13.4 Check the single-use dispersive electrode for flaws, damage, discoloration, lack of adhe- siveness, and dryness before application, and do not use the electrode if any of these con- ditions are present.49 [Recommendation] Using a single-use dispersive electrode that has any of these problems increases the poten- tial for patient injury, especially at the disper- sive electrode site.49 A clinical guideline supports checking the dispersive electrode for the issues listed.86 3.14 When placing a single-use dispersive electrode on the patient, • • place the electrode • after final patient positioning69; • close to the surgical site25,38,49,69,110,111; • on clean, dry skin20,69; • over a large, well-perfused muscle mass25,38,44,49,69,110; • on the surgical side when indicated (eg, left or right shoulder, hip, kidney)25,49,110; • to avoid any metal or monitoring leads between the dispersive and active elec- trodes49,110,111; • away from a warming device111; and • so that it is in uniform contact with the patient’s skin2,20,40,44,49,112; • • do not place the electrode • over a bony prominence or potential pres- sure point20,25,38,44,49,69,110; • over scar tissue20,25,38,40,44,49,69; • over an implant containing metal com- ponents25,38,44,49,69; • over skin folds38; • on areas with dense hair20,38,49,69; • on areas where pooling of prep solutions may occur20,44,46; • where the cord could cause a pressure injury; or • on areas distal to tourniquets69; • • do not reposition or reapply the electrode49; and • • do not alter (eg, cut, fold) the electrode. [Recommendation] Clinical guidelines support implementing the cri- teria listed.51,59,86 A lack of good contact between the skin and the dispersive electrode can result in patient injuries.10 Skin burns have been reported at the dispersive elec- trode site.2,39,47,48,88,89,112,113 Nguyen et al53 conducted a bench study to exam- ine the effect of monopolar radio-frequency energy on the temperature of bovine myocardial tissue sur- rounding implanted metal objects. The researchers found the temperature was elevated in the tissue surrounding the objects by a mean of 5.8° C to 8.9° C (10.4° F to 16° F) when the objects were between the active and dispersive electrodes and at a distance of 1 mm from the site of the active electrode. The researchers concluded that the dispersive electrode should be placed in a location with no metal objects between it and the active electrode. 3.15 Check the single-use dispersive electrode for uniform contact if the patient has been moved or repositioned or if any tension has been applied to the dispersive electrode cord.49,88,112 [Recommendation] Injuries have been reported that resulted from inadequate adhesion of the dispersive electrode.49,88,112 3.15.1 Perform corrective measures if inadequate contact between the single-use dispersive electrode and the patient occurs, including • • removing oil, lotion, moisture, or prep solution86,88; • • removing excessive hair86,88; and • • applying a new single-use dispersive elec- trode to the same or a different site.49,86 [Recommendation] Performing these interventions assists with obtaining adequate adhesion of the dispersive electrode.49,88,112 3.16 When using a dual-foil, single-use disper- sive electrode, use a generator with return- electrode contact quality monitoring.2,25,29,48,49,114 [Recommendation] Return-electrode contact quality monitor- ing confirms that there is adequate contact between the return electrode and the patient and inhibits the output of the ESU if the dis- persive electrode contact area is too small. An audible alarm or visual indicator alerts the user to a misconnection.2,25,48,49,114 A clinical guideline supports using a dual single- use dispersive electrode with a generator that has return-electrode contact quality monitoring.86 3.17 When using a reusable dispersive electrode cable, • • securely connect the cable to the electrode, • • cover the gel-free connecting lugs or tabs with the attachment clip, and • • inspect the cable for wear and replace it when worn.49,86 [Recommendation] Connecting the cable to the electrode in this fash- ion helps prevent a patient burn related to the elec- trical current traveling from the exposed lugs or tabs to the patient.49 I-04_ElectroSafety-25_110124.indd 153 I-04_ElectroSafety-25_110124.indd 153 11/7/2024 10:34:33 AM 11/7/2024 10:34:33 AM

154 ELECTROSURGICAL SAFETY 3.18 When using and storing a reusable dispersive electrode, follow the manufacturer’s IFU.86 [Recommendation] 3.18.1 Select the size based on the patient’s weight, as described in the manufacturer’s IFU (eg, neonate, infant, pediatric, adult).86 [Recommendation] P 3.18.2 Confirm the manufacturer’s beyond-use date has not passed prior to application, and do not use the electrode if the beyond-use date has passed.86 [Recommendation] 3.18.3 Use the reusable dispersive electrode with minimal or no material between the elec- trode and the patient. [Recommendation] 3.18.4 Check the reusable dispersive electrode for tears or breaks in the surface material before and after use, and do not use the electrode if damage is present.46,86 [Recommendation] 3.18.5 Clean and disinfect the reusable dispersive electrode with a health care facility– approved and Environmental Protection Agency–registered agent between uses in accordance with the manufacturer’s IFU. [Recommendation] 3.18.6 Inspect the reusable, dispersive electrode cables for integrity before use, and replace the cables if damaged.46 [Recommendation] 3.19 When high current (eg, ablation) is used for a prolonged period and there are no specific manufacturer’s instructions, either a second single-use dispersive electrode or a reusable dispersive electrode may be used.112,113 [Conditional Recommendation] Using a second single-use dispersive electrode site helps increase the dispersive area, therefore decreasing the potential for a burn when high- current, long-activation-time procedures are per- formed.112,113 Reusable dispersive electrodes cover a large dispersive area. 4. Minimally Invasive Surgery 4.1 Use conductive trocar systems when using elec- trosurgery during minimally invasive sur- gery.1,20,24,46,48,99 [Recommendation] Conductive trocar cannulas provide a means for the electrosurgical current to flow safely between the cannula and the abdominal wall. This reduces high-density current concentration and heating of non-target tissue, therefore decreasing the potential for injury related to capacitive coupling.1,24,46,99 Clinical guidelines support the use of conductive trocar systems.59,86 4.2 Prevent contact between the monopolar active electrode and other conductive instruments or materials.1,29,40,69,115 [Recommendation] Preventing contact between the monopolar active electrode and other conductive instruments or materials helps decrease the potential for a burn related to direct coupling.1,29,40,69,115 A clinical guideline supports preventing contact.86 4.3 Use an active electrode monitoring and shield- ing device.116-118 [Recommendation] Moderate-quality evidence116-118 and clinical guide- lines59,86 support using an active electrode monitor- ing and shielding device. In a laboratory study, Martin et al116 measured the amount of energy transferred via capacitive coupling to determine whether the use of active electrode monitoring equipment decreased the amount of energy that escapes from a laparoscopic active elec- trode with insulation defects during laparoscopic sur- gery. The researchers used porcine tissue as a substi- tute for human tissue. They found that the use of the active electrode monitoring equipment decreased the number of tissue burns and the amount of energy transferred via capacitive coupling. Guzman et al117 reviewed 192,794 patients’ records and found that 694 patients had experienced an acci- dental puncture or laceration during laparoscopic abdominal procedures. The researchers recom- mended using an active electrode monitoring and shielding device to decrease the incidence of adverse events. Mendez-Probst et al118 conducted a descriptive laboratory study to identify the amount of current leakage in 37 robotic instruments with visually intact insulation. Stray currents were found from 100% of the instruments with a mean leakage of 4.1 W. The researchers recommended adoption of an active electrode monitoring technology program. 5. Implanted Electronic Devices 5.1 When use of electrosurgery is possible for a patient with an IED, the anesthesia professional and the perioperative RN should consult with the team managing the IED preoperatively to define interventions necessary for safe management of the device during the intraoperative and postop- erative phases of care.30,31,51 [Recommendation] P I-04_ElectroSafety-25_110124.indd 154 I-04_ElectroSafety-25_110124.indd 154 11/7/2024 10:34:33 AM 11/7/2024 10:34:33 AM

ELECTROSURGICAL SAFETY 155 Preoperatively consulting with the team manag- ing the IED helps the perioperative team obtain directions for caring for the patient.25,30,31,37,38,44,51,111,119-132 Team members that the anesthesia professional and the perioperative RN may contact are listed below for specific devices: • • Pacemaker—cardiologist, electrophysiology team, health care industry representative • • Cochlear implant—implanting surgeon, health care industry representative • • Neurostimulator device—implanting surgeon, neurologist, health care industry representative • • Implanted infusion pump—implanting surgeon, physician responsible for managing care, health care industry representative 5.2 In an emergent situation when the team managing the IED cannot be contacted, the anesthesia profes- sional and the preoperative RN or the RN circulator should consult with the implant manufacturer to determine interventions to perform for safe use of electrosurgery.31,37,133 [Recommendation] Many IED manufacturers have a 24-hour hotline available to assist in caring for a patient with an IED.37 Clinical guidelines support contacting the manufacturer.31,59,133 5.2.1 Provide the following information to the team managing the device or the IED manu- facturer’s representative: • • the patient’s level of dependence on the IED; • • the availability of a person competent to perform reprogramming if applicable; • • the scheduled procedure, including the site; • • whether cardioversion or defibrillation is anticipated; • • the intraoperative patient position; • • the potential for EMI based on the type of electrosurgical device to be used (eg, radio-frequency ablation, monopolar, bipolar); • • other potential sources of EMI (eg, shiver- ing, nerve stimulators, large tidal volumes, extracorporeal shockwave lithotripsy); • • the location of the generator; • • the procedure room location (eg, OR, interventional radiology); and • • the postprocedure patient disposition (eg, outpatient, inpatient).30,122,124 [Recommendation] Clinical guidelines support providing this information.30,31,133,134 Obtaining this information will help the team managing the patient with an IED to determine the interventions to be per- formed during the perioperative period.30,122,124 The surgical site may determine the actions required. For example, if the site is 15 cm away from the generator and the dispersive electrode is placed so the generator is not in the pathway between the active and dispersive electrode, then no interventions beyond monitoring may be required.134 5.2.2 A chest x-ray may be performed if identifi- cation of a cardiac implanted electronic device (CIED) is not possible.120,128 [Conditional Recommendation] Performing a chest x-ray helps confirm the device lead placement and identify the device. The device identification is visible on a chest x-ray.120,128 5.3 Create an interdisciplinary team to develop and implement a clinical support tool that contains interventions to perform for a patient with an IED, for use when the team managing the IED or the manufacturer cannot be contacted. [Recommendation] A clinical support tool can provide instructions for caring for the patient with an IED in an emer- gency situation or when the necessary people are not available to provide direction. 5.3.1 Include the following personnel on an inter- disciplinary team responsible for developing the clinical support tool: • • perioperative RNs, • • scrub personnel, • • anesthesia professionals, • • surgeons who implant and manage IEDs, • • representatives from IED manufacturers, and • • other stakeholders. [Recommendation] Having a diverse team helps cover all aspects of patient care. 5.3.2 The IED clinical support tool should contain interventions specific to each IED (eg, repro- gramming, removal of external parts, use of alternatives to monopolar electrosurgery, when to inactivate the device). [Recommendation] 5.3.3 Determine the applicable intervention based on the following IED information: • • type of implanted device; • • patient’s level of dependence and severity of symptoms if the device is turned off; • • location of the device and leads (eg, within or outside the path between the active and dispersive electrodes); • • device manufacturer and model; • • clinical indication for the device; I-04_ElectroSafety-25_110124.indd 155 I-04_ElectroSafety-25_110124.indd 155 11/7/2024 10:34:33 AM 11/7/2024 10:34:33 AM

156 ELECTROSURGICAL SAFETY • • date of the last device interrogation or evaluation; • • battery life; • • device function; and • • device settings. Obtain the following additional information when the patient has a CIED: • • lead polarity (eg, unipolar or bipolar), • • need for programming, • • response to a magnet, • • presence of an alert status on the genera- tor or on the lead, and • • the last pacing threshold.30,122,124 [Recommendation] Clinical guidelines support using the listed information when determining interventions to be performed for the patient with a CIED.30,31,133 Gath- ering the listed information helps the team deter- mine the necessary interventions to perform on the patient with an IED.37,49,119-125,127,129,130,135,136 The information may be available from the patient assessment, the patient’s caregivers, the implant identification (ID) card carried by the patient, and the medical record. 5.4 Perform the following interventions when a patient has an IED: • • inactivate or reprogram the IED if it is safe to do so25,31,124,131,132,136,137; • • place the dispersive electrode as close as possible to the surgical site, but as far as possible from the device’s generator and leads49,124,125,131,132,136,137; • • place the active electrode cord away from the pulse generator25,131; • • verify that the implanted device and leads are not between the active electrode and the dis- persive electrode49,124,125,131; • • activate the ESU for the shortest amount of time possible25,49,124,131,132,136; • • use the lowest effective power settings124,131; and • • use alternative technology (eg, electrocau- tery, bipolar forceps, ultrasonic technol- ogy, a ferromagnetic surgical system) instead of monopolar electrosurgery when possible.25,49,124,125,132,136,137 [Recommendation] P Low-quality evidence25,49,124,125,131,132,136-140 and clinical guidelines30,31,59,134 support performing the listed interventions to decrease the risk of adverse events in patients with IEDs. Bipolar electrosurgery does not produce EMI unless the forceps come into direct contact with the CIED.86 Short activation bursts and a low power set- ting are beneficial because the amount of EMI pro- duced is directly proportional to the amount of time of activation and the power setting.132 The greater the distance between the generator and the active and dispersive electrode, the smaller the amount of EMI.132 In a nonexperimental study, Friedman et al141 reviewed the records of 103 patients who had sur- gery after cardioverter defibrillator implantation. The researchers found no EMI with the use of bipolar technology but 11 cases of EMI with the use of mono- polar technology. The researchers also found no cases of EMI when the surgical site and the disper- sive electrode site were both below the hips. The researchers concluded that EMI does not occur when bipolar technology is used or when the surgical site and the dispersive electrode site are below the hip joint. In a cadaveric study, Jeyakumar et al142 assessed whether changes to an implanted cochlear device occurred during the use of monopolar electrosur- gery and if there were changes in the intracochlear temperature. The researchers found no changes in the device, and the temperature of the cochlear device did not increase after 30 minutes of electro- surgical energy delivered at 30 W into the oral cav- ity. The researchers concluded that no damage or temperature increase occurred, but further study is required before changing existing practice. 5.5 Perform the following interventions when a patient presents with a CIED: • • place a magnet over the pacemaker to repro- gram the device when indicated,2,25,36-38,49,119, 121,122,124,126,128,129,135,140,143,144 • • use an electrocardiogram system that has five or more leads,30,37,127,140 • • use a beat-to-beat indicator (eg, arterial line, pulse oximeter),2,37,120-122,126,129,143 • • have temporary pacing equipment and an exter- nal defibrillator immediately available,25,31,36,37,49,121 • • have a magnet immediately available if the pacemaker will respond to a magnet and one is not used for reprogramming2,124,140, and • • use continuous cardiac monitoring whenever the CIED has been reprogrammed.2,25,35,44,49,69,120,122, 123,126,128,129,135,144 [Recommendation] Low-quality evidence2,25,33,35-38,44,49,52,69,120-125,127-130,136,140,141,144,145 and clinical guidelines30,31,51,133,134 support performing the listed precautions to decrease the risk of adverse events in a patient with a CIED. Use of a magnet is not always indicated because the response of a pacemaker is based on the model, age, and manufacturer of the device. In addition, use of a magnet on a pacemaker with a low battery may lead to device failure because the power of the bat- tery may become rapidly exhausted.25,37,49,119,121,126,135 Use of a magnet on a pacemaker will switch it to an I-04_ElectroSafety-25_110124.indd 156 I-04_ElectroSafety-25_110124.indd 156 11/7/2024 10:34:33 AM 11/7/2024 10:34:33 AM

ELECTROSURGICAL SAFETY 157 asynchronous mode and inactivate the rate response feature. Use of a magnet on an implantable cardio- verter defibrillator inactivates the anti-tachycardia features and does not alter the pacing mode.134 Pacemakers may incorrectly interpret the elec- trical signals from an electrosurgical device, lead- ing to a misfiring or failure to fire that can result in arrhythmia, such as ventricular fibrillation or ven- tricular tachycardia.49 The tolerance for EMI is different for different types of pacemakers (unipolar, bipolar) and differ- ent makes and models of pacemakers.37 Use of a five-lead electrocardiogram configura- tion provides multiple leads to view in case one lead is obscured or difficult to interpret because of EMI.37 Electromagnetic interference is unlikely with a CIED generator implanted in the thoracic area when the procedure is below the umbilicus and the return electrode is placed on the lower body (thigh or gluteal area).31,140 Using a beat-to-beat indicator provides a signal for the pulse in case the electrocardiogram is obscured or difficult to interpret because of EMI.2,37 Moderate-quality evidence shows that EMI affects pacemakers.32,84,85,145 Lin et al32 found four patients (0.3% of 1,398 patient records reviewed) who developed output failure or an inappropriate low pacing rate during use of electrosurgery. The patients were undergo- ing pacemaker generator replacement or upgrade surgeries. The researchers also reviewed the MAUDE database for events or patient injuries related to pacemaker malfunction and found 37 cases of pacemaker malfunction related to the use of electrosurgery. The adverse events included the pacemaker resorting to the backup pacing mode (32.4%), loss of output/capture (59.5%), inappropri- ate low pacing rate (5.4%), and ventricular fibrilla- tion (2.7%). The dates of review of the MAUDE data- base were not included in the report. The researchers concluded that certain pacemaker models may exhibit loss of pacing or inappropriate low pacing rate during the use of monopolar elec- trosurgery during pacemaker replacement or upgrade surgery. Paniccia et al145 measured the average maximum EMI that occurred during activations of various sur- gical energy-based devices on a cardiac device implanted in a pig. The researchers found the mean EMI transferred to the implanted cardiac device varied for each device: • • 0.1 mV for a traditional bipolar device at 30 W, • • 0.004 mV for an advanced bipolar device, • • 0.1 mV for ultrasonic shears, • • 0.5 mV for a monopolar device at 30 W coagu- lation, • • 0.92 mV for a monopolar device at 30 W blend, • • 0.21 mV for a monopolar device without a dis- persive electrode, • • 3.48 mV for a plasma energy-generating device, and • • 2.58 mV for an argon beam coagulator. The researchers concluded that bipolar and ultrasonic devices are useful for decreasing the EMI on implanted cardiac defibrillators. Govekar et al85 conducted a laboratory study to determine whether the location of the dispersive electrode affected the functionality of a pacemaker in the heart of a pig. The dispersive electrode was placed on the right gluteus, left gluteus, right shoulder, and left shoulder. The ESU was set at a power of 30 W in the coagulate mode. The active electrode was activated for 5 seconds at a distance of 7.5 cm from the pacemaker generator. The researchers found more beats were dropped (1.5 versus 0.2) when the current vector traveled from the active electrode to the dispersive electrode through the pacemaker or through the leads. The researchers concluded that the dispersive electrode should be located such that the current vector does not pass through the pacemaker or the leads. In a laboratory study, Robinson et al84 examined the effect of monopolar electrosurgery devices on CIEDs implanted in pigs. The researchers mea- sured the amount of EMI present after they changed the orientation of the active electrode cord from across the chest wall to coming from the feet, which changed the current vector so it did not flow across the CIED. The researchers found significantly less EMI when the orientation of the active electrode was from the feet and the current vector avoided the CIED. They concluded that the active electrode cord should be placed at the greatest distance possible from the CIED gen- erator, and the CIED generator should not be in the pathway of the current from the active elec- trode to the dispersive electrode. 5.6 Take the following precautions for a patient with an existing cochlear implant: • • remove all external components, • • do not use electrosurgical devices including bipolar devices within 2 cm to 3 cm of the implant generator electrodes,25,123,146,147 • • use only bipolar electrosurgery above the clavicles,147 and • • monopolar electrosurgery may be used below the clavicles if the dispersive electrode is also below the clavicles.147 [Recommendation] P I-04_ElectroSafety-25_110124.indd 157 I-04_ElectroSafety-25_110124.indd 157 11/7/2024 10:34:33 AM 11/7/2024 10:34:33 AM

158 ELECTROSURGICAL SAFETY Keeping the recommended distance from the device and removing external parts reduces the risk of accidental activation or damage to the device.25,123,146,147 5.7 Decrease the amplitude to the lowest level and then inactivate deep brain stimulators, sacral nerve stimulators, spinal cord stimulators, vagal nerve stimulators, or gastric pacemakers, if this is advised by the manufacturer and can be toler- ated by the patient.124,125,137 [Recommendation] Low-quality evidence124,125,137 and a clinical guide- line30 support inactivating the device to prevent inadvertent stimulation and decreasing the ampli- tude to decrease the amount of stimulation if the device is accidently turned on. 5.8 Consult with the team managing the implant to determine postoperative interventions as soon as possible after surgery.30 [Recommendation] Consultation allows the team managing the implant to confirm functioning of the implant or to restart the implant if it has been inactivated.37,121,124-126,128,132,135,139 Clinical guidelines support postoperative contact with the team managing the implant.30,31 5.9 Provide education and postoperative instructions to patients and their caregivers on the effects of electrosurgery on IEDs.148 [Recommendation] P Frampton and Mitchell148 conducted a survey of 50 adults with cochlear implants and the parents of 50 children with cochlear implants to determine their knowledge of the risks for complications when a cochlear implant is present and electrosurgery is used during head and neck procedures. The researchers found that 86% of the adult patients pos- sessed an implant ID card, and 71% carried it with them. Seventy-seven percent stated that if surgery were required, they would show their implant ID card and discuss the issue of electrosurgery with the physician. Eighty-four percent of the parents reported that the child possessed an ID card; 8% of parents ensured the child carried the card at all times, and 12% reported that they carried the card for the child. Seventy-six percent of the parents believed that sur- geons would be unaware of the electrosurgical restrictions, but none of the 12% of parents whose children had undergone